Getting acquainted
AC and DC (Battery) Operation
1-12
is to be stored for extended periods (more than one month) without AC power. Note
on the instrument that the battery has been removed. After an extended storage
period, test the battery according to the battery capacity check as described in
Maintaining the Battery on page 8-3.
This battery was selected because it provides optimum performance and battery life
over a wide range of operating conditions. The life expectancy of this battery is
dependent on many variables, including temperature and usage. Periodically check
the battery capacity to determine whether to replace it. The battery capacity check is
described in Maintaining the Battery on page 8-3.
When plugged into AC power, the CodeMaster XL+ will function normally with no
battery installed, however the time required to charge the defibrillator will increase.
If the CodeMaster XL+ is operated without a battery installed,
clearly mark the instrument that it does not have a battery and
requires AC power to operate. When a CodeMaster XL+ has no
battery installed and is plugged into AC power, the front panel
light will be on and the light will be off.
Properly dispose of or recycle depleted batteries according to
local regulations. Do not disassemble, puncture or incinerate the
disposed batteries.
NOTE
WARNING
WARNING