Troubleshooting
Troubleshooting
7-3
If the Defibrillator does not Charge
If the defibrillator does not charge, perform the following steps.
1 Verify the proper setting of the Energy Select control.
2 If it is correct, follow these steps.
a. Turn the Energy Select control to Off (Standby), and then back to the
desired energy setting.
b. Press again.
3 If the instrument remains unable to charge, turn the Energy Select control to Off
(Standby) and use a backup defibrillator.
4 Press to print an event summary and keep any ECG strips from the
defibrillator for later evaluation.
5 Alert appropriate service personnel.
If the Defibrillator does not Deliver a Shock
If the defibrillator does not deliver a shock, perform the following steps.
DEFIB DIS-
ARMED
CodeMaster XL+ has internally dis-
charged energy and is now dis-
armed.
CHECK
RECORDER
The recorder door is open.
or
The recorder is out of paper.
Close the recorder door or
replace the paper roll as
described in “Changing the
Recorder Paper” (page 7-1).
Table 7-2 Defibrillator Messages
Message Cause Possible Solutions
Charg
Revie