A-14
Refractory Period: 40 to 80 ppm 340 msec; 90 to 180 ppm 240 msec.
SpO
2
(Optional)
Measurement Range: 0 to 100%.
Accuracy with HP M1190A sensor: 1 standard deviation, 65-80%: ±2.5%, 80-
100%: ±1.5%, resolution: 1%.
Averaging: 8 beats.
SpO
2
alarm limits — range: 100/90, 100/85, 100/80.
SpO
2
alarm delay: ten seconds after value drops below the low alarm setting.
INOP alerts: Triggered by disconnected sensor, noisy signal, light interference or
low signal.
Pulse rate measurement range: 30 to 300 bpm ± 1%; resolution 1 bpm.
Pulse amplitude indicator: Indicates pulsatile activity.