Section V
Table 5-3 and Figure 5-I
Model 4342A
Table 5-3. Frequency Accuracy Check.
Frequency
Frequency Measured
Range
Dial Setting Accuracy
22k - 70k
2.2
L
5.0
7.0
*l. 5%
il. 0%
il. 5%
&l. 5%
70k - 220k
I
220k - 700k
700k - 2.2M
2.2M - 7M
7M - 22M
22M - 70M
7.0
il. 5% 68.950 - 71.050 kHz
L *l. 0%
78.822 - 80.413
kHz
15
il. 5% 147.75 - 152.25 kHi
22
il. 5% 216.70 - 223.30 kHz
Counter Reading
21.670 - 22.330 kHz
24.922 - 25.424
kHz
49.250 - 50.750 kHz
68.950 - 71.050
kHz
216.70 - 223.30 kHz
249.22 - 254.24 kHz
492.50 - 507.50 kHz
689.50 - 710.50 kHz
689.50 - 710.50 kHz
788.22 - 804.13 kHz
1477.5 - 1522.5 kHz
2167.0 - 2233.0
kHz
2167.0 - 2233.0 kHz
2492.2 - 2542.4
kHz
4925.0 - 5075.0 kHz
6895.0 - 7105.0 kHz
6895.0 - 7105.0 kHz
7882.2 - 8041.3 kHz
14.775 - 15.225
MHz
21.670 - 22.330 MHz
21.560 - 22.440 MHz
24.922 - 25.424 MHz
49.000 - 51.000 MHz
68.600 - 71.400 MHz
-yl
4342A
r
DIGITAL VOLTMETER
@1@
JpqL!J
(REAR PANEL)
Figure 5-l. Q Range Check.