EasyManua.ls Logo

HYDAC ELECTRONIC ETS 300 - 11 Technical Specifications

HYDAC ELECTRONIC ETS 300
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ETS 300 13
Status: 29.01.2014 HYDAC ELECTRONIC GMBH Part No.: 669651 E
11 Technical specifications
Input data for built-in senso
r
Measuring range -25 .. 100 °C (-13 .. 212 °F)
Probe length 18 mm
Pressure resistance 600 bar
Mechanical connection G1/2 A DIN 3852
Torque value 45 Nm
Parts in contact with medium Mech. conn.: stainless steel
Seal: FPM
Input data for separate senso
r
Measuring range
1)
-30 .. 150 °C (-22 .. 302 °F)
Connection, separate temp. probe Female cable connection M12x1, 4-pole
Output data
Accuracy (display, analogue output) ± 1.0 °C ( ± 2.0 °F)
Temperature drift (environment) ± 0.015 % FS / °C max. zero point
± 0.015 % FS / °C max. range
Analogue output (optional)
Signal 4 .. 20 mA, ohmic resistance 400
Corresponds to
-25 .. +100 °C with built-in sensor
-30 .. +150 °C with separate sensor
Switching outputs
Type PNP transistor switching outputs
Switching current max. 1.2 A per output
Switching cycles > 100 million
Rise time acc. to DIN EN 60751 t
50
= 3s / t
90
= 9s
Ambient conditions
Ambient temperature range -25 .. +80 °C
Storage temperature range -40 .. +80 °C
Fluid temperature range
2)
for the probe (for built-in sensor)
-40 .. +100 °C / -25 .. +100 °C
-mark EN 61000-6-1 / 2 / 3 / 4
Vibration resistance to
DIN EN 60068-2-6 (0 .. 500 Hz)
10 g
Shock resistance to
DIN EN 60068-2-29 (1 ms)
50 g
Protection class to IEC 60529 IP 65
Other data
Supply voltage 20 .. 32 V DC
Current consumption approx. 100 mA without switch output
Residual ripple of supply voltage 5 %
Display 3-digit, LED, 7 segment, red,
height of digits 9.2 mm
Weight 300 g
Note.: Reverse polarity protection of the supply voltage, excess voltage, override, short circuit
protection are provided. FS (Full Scale) = relative to the complete measuring range
1) Depending on the temperature range of the temperature sensor fitted, this can reduce the
display range of the ETS 300 with a separate sensor input.
2) -25 °C with FPM seal, -40 °C on request.