8.5-9
Observe the vacuum gauge and ensure the gauge displays a minimum 0.7 Bar.
Figure 8.5-14
If the value remains stable and is above 0.7 Bar press the ‘Yes’ button.
PLEASE NOTE: If the vacuum sealing plate is incorrectly fitted, the pressure will drop
(observe vacuum gauge). In this event, check the seal between the printhead and Vacuum
Sealing Plate and/or pipe connections to the Priming Ink Reservoir.
If a ‘good’ seal is not established the priming procedure will fail.
Once a good seal is established and the vacuum gauge remains stable, press ‘Continue’ on
CLARiTY™ to restart the pump and the test will be repeated.