EasyManua.ls Logo

IFM Electronic SM2000 - 10�6�7 Setting the Error Behaviour of the Outputs; 10�6�8 Configuring the Empty Pipe Detection as Diagnostic Output; 10�6�9 Activating; Deactivating Empty Pipe Detection; 10�6�10 Time-Delay Empty Pipe Detection

IFM Electronic SM2000
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
UK
10.6.7 Setting the error behaviour of the outputs
► Select [FOU1] and set the value:
1� Switching output:
- [On] = output 1 switches ON in case of an error�
- [OFF] = output 1 switches OFF in case of an error�
- [OU1] = output 1 switches irrespective of the error as defined with the
parameters�
2� Frequency output:
- [On] = 130% of FrEP
- [OFF] = 0 Hz
- [OU1] = continues running
► Select [FOU2] and set the value:
- [On] = output 2 switches ON in case of an error, the analogue signal
goes to the upper error value�
- [OFF] = output 2 switches OFF in case of an error, the analogue signal
goes to the lower error value�
- [OU2] = output 2 switches irrespective of the error as defined with the
parameters� The analogue signal corresponds to the measured value�
10.6.8 Configuring the empty pipe detection as diagnostic output
► Select [OU1] or [OU2] and set [dOU]�
► Select [P-n] and set [PnP] or [nPn]�
The empty pipe detection is only effective if it is activated at [EP�
On]→10.6.9.Whentheemptypipestateisdetected,thediagnostic
output is inactive�
10.6.9 Activating / deactivating empty pipe detection
► Select [EP�On] and set the function:
- [OFF] = empty pipe detection deactivated�
- [On] = empty pipe detection activated�
10.6.10 Time-delay empty pipe detection
► Select [dEP�E] and set the delay time from 0…30 s, at which the signal
should be provided when the pipe is empty�
► Select [dEP�F] and set the delay time from 0…30 s, at which the signal
should be provided when the pipe is full�

Table of Contents