MDtest (100-140 V) - User’s Manual 4 Measurements
Procedure:
• Press the [5 IDIR] key and release it.
• Press the [Rever] key repetitively in order to measure with L-N reversal. The
symbol L-N reversed is displayed.
• Connect the tested appliance: plug in its mains plug into Test Socket 1 and
plug the applied part into the PAT terminal. If necessary (e.g. protective
class II appliances), plug the test lead into the PE terminal.
• Press the [TEST] key.
• Switch on the tested appliance’s mains switch.
• Press the test tip against the exposed conductive part of the tested appliance.
• The HOLD function can be used (press the [H] key).
• Switch off the tested appliance’s mains switch.
• Finish the measurement (see chapter 4.2.1).