EasyManuals Logo

IMC C SERIES User Manual

IMC C SERIES
185 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
© 2018 imc Test & Measurement GmbH
imc C-SERIES - Manual, Version 4 R 3 - 2018-10-19
158 Technical Specs
6.10.2 ACC/DSUB-ICP
Parameter
Value (min / max)
Remarks
option for
C-10xx, C-12xx, C-41xx, C-50xx, C-60xx,
C-70xx
Inputs
4
2
differential, not isolated
ACC/DSUB-ICP4
ACC/DSUB-ICP2
Input coupling
DC
ICP
current source, 1st order high-pass
Current drain per connector
<0.2 A
<0.1 A
ACC/DSUB-ICP4
ACC/DSUB-ICP2
Voltage measurement
Input voltage max.
voltage
ICP
±60 V
-3 V to 50 V
±3 V
permanent to chassis
at +IN1, ..., +IN2 bzw. +IN4
at -IN1, ..., -IN2 bzw. +IN4
Input impedance
voltage
ICP
1 M
10 M
20 M
0.33 M
0.91 M
depending on the measurement ranges
of the measurement inputs
differential
single end
ICP™-, DELTATRON®-, PIEZOTRON®-Sensors
Highpass cutoff frequency
3 Hz
1 Hz
±20 %
±20 %
-3 dB, AC, corresponding to input
impedance of the used measurement
input
1 M
10 M, 20 M
ICP-current source
4.2 mA
±10%
Voltage swing
25 V
>24 V
Source impedance
280 k
>100 k
Find here the description of the IEPE (ICP)-expansion plug .