9 | Cloud Hunting (Portable Monitor) Mode INFICON
16 / 394 074-708-P16A D-TEK Stratus Operating Manual
9 Cloud Hunting (Portable Monitor) Mode
ppm reading
max reading
side bar
Cloud Hunting
symbol
Cloud Hunting
mode is identified by a large ppm reading on the display and the words
Cloud Hunting
on the bottom, along with a cloud symbol. The side bar increases and
decreases with changes in the ppm reading.
There is no sensitivity setting in Cloud Hunting mode.
1.
Slowly move through the suspect areas and observe the ppm reading.
2.
Follow the ppm reading to find areas of higher refrigerant concentration. The
higher the number, the higher the concentration.
3.
Press the
SENS/ZERO
button to enable and disable the
MAX
feature. When
enabled, the highest ppm level observed is shown below the main ppm display.
To reset the
MAX
reading, long-press the
SENS/ZERO
button or toggle the
feature off and back on again.
D-TEK Stratus uses a patent-pending switching valve in
Cloud Hunting
mode to
constantly compare the sample from the tip of the probe with the air inside the body of
the leak detector (the reference sample). This technology is what allows D-TEK
Stratus to work without the use of a carbon filter. Lingering for several minutes in an
area with a high concentration of refrigerant may cause the reference sample to
English
English

Table of Contents

Other manuals for Inficon D-TEK Stratus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inficon D-TEK Stratus and is the answer not in the manual?

Inficon D-TEK Stratus Specifications

General IconGeneral
BrandInficon
ModelD-TEK Stratus
CategorySecurity Sensors
LanguageEnglish

Related product manuals