16 | Extra-long Probe INFICON
24 / 394 074-708-P16A D-TEK Stratus Operating Manual
16 Extra-long Probe
D-TEK Stratus includes an extra-long replacement probe for leak checking in hard-to-
reach areas.
To install the extra-long probe:
1.
Unscrew the standard probe from the D-TEK Stratus body using a 10 mm
wrench.
2.
Screw on the extra-long probe to approximately 35 in·lb (4 N·m). Do not
overtighten.
3.
Unscrew the probe tip from the standard probe and remove the filter (or use a
new one).
4.
Insert the filter into the extra-long probe.
5.
Screw the probe tip onto the extra-long probe. Do not overtighten.