14 Removing and Installing the Sensor
D-TEK Stratus uses a cartridge style sensor that is quick and easy to remove and
replace in the field. In addition to the standard sensor, INFICON offers a CO
2
specific
sensor for use in refrigeration and air conditioning applications. See Replacement
Parts and Accessories [}28].
eject button
sensor release
tab
To replace the sensor:
1.
Press the eject button on the back of D-TEK Stratus and remove the battery
door.
2.
Grasp the sensor release tab and gently pull it out.
3.
Align the new sensor with the rails.
4.
Gently push the sensor along the rails until it is fully inserted.
ð Do not force the sensor. If it does not slide freely, check the alignment and try
again
5.
Reinstall the battery door.