INFICON Manual Zero Mode | 11
074-708-P16A D-TEK Stratus Operating Manual 19 / 394
11 Manual Zero Mode
leak indicator
(showing full
alarm)
Manual Zero
symbol
Manual Zero
mode looks and operates similar to
Pinpoint
mode, but is identified by
the text
Manual
Zero
and the
Manual
Zero
symbol on the bottom bar.
Manual
Zero
mode allows the user to manually zero to the background refrigerant by pressing the
SENS/ZERO
button. Once the new zero point is set, D-TEK Stratus will not alarm
unless a higher concentration of refrigerant is detected.
Manual Zero
mode beeps faster at the zero point than other modes. If the
concentration is lower than the current zero point, the beeping slows. This allows the
user to know if they are moving away from the leak by listening to a change in the
beep rate.
There is no sensitivity setting in
Manual Zero
mode.
Manual Zero
mode requires an extra warm-up time of up to 15 minutes for optimal
performance.
English
English