7 | Warning and error messages INFICON
80 / 108 XL3000flex-Operating-instructions-jina83en1-07-(2105)
Warning
(Wrn)
Error (Err)
Error message
LDS3000
Error number Limit values Cause
LDS1000
Protocol
Binary or
ASCII
protocol
compatibilit
y mode
LDS1000/
LDS2010
Wrn125 I/O module not connected 99 99 Connection to I/O module
interrupted
Wrn127 Wrong bootloader version 99 99 Boot loader not compatible with
application
Err129 Incorrect device
(EEPROM)
99 99 EEPROM does not contain any
compatible data
Err130 Sniffer not connected 99 99 The sniffer line is not connected.
Wrn132 SL3000 not supported Only the SL3000XL may be used
with the XL3000flex
Wrn150 Pressure sensor 2 is not
connected
Connecting pressure sensor
PSG500 to a FINE connection.
2xx operating voltage error
Wrn201 U24_MSB too low 24 120 21.6V 24V power supply pack
Wrn202 U24_MSB too high 24 120 26.4V 24V power supply pack
Wrn203 24V_PWR12 voltage out of
range (TL_valve/GB_valve)
24 120 20V
30V
Short circuit at valve 1 (calibrated
leak) or valve 2 (gas ballast)
Wrn204 24V_PWR34 voltage out of
range
(valve 3/4)
24 120 20V
30V
Short circuit at valve 3 or valve 4
Wrn205 24V_PWR56 voltage out of
range (Sniff_valve/valve6)
24 120 20V
30V
Short circuit at valve 5 (sniff) or
valve 6
Wrn221 Internal voltage 24V_RC
voltage out of range
24 120 20V
30V
Short circuit 24V at the control unit
output
Wrn222 Internal voltage 24V_IO
voltage out of range
24 120 20V
30V
Short circuit 24V at IO output
Wrn223 Internal voltage 24V_TMP
voltage out of range
24 120 20V
30V
Short circuit 24V of the TMP
Wrn224 Internal voltage 24V_1
(Pirani) voltage out of
range
24 120 20V
30V
Short circuit 24V
Pressure sensor PSG500 (1,2,3),
sniffer line