LDS3000
Error number Limit values Cause
LDS1000
Protocol
Binary or
ASCII
protocol
compatibilit
y mode
LDS1000/
LDS2010
Wrn240 Voltage +15V out of range 24 120 +15V too low, IF board or MSB
defective
Wrn241 Voltage -15V out of range 24 120 -15V too low, short circuit at
preamplifier, IF board or MSB
defective
Err242 +15V or -15V voltage
shorted
24 120 +15V or -15V too low, short circuit
at preamplifier, IF board or MSB
defective
Wrn250 REF5V voltage out of
range
24 120 4.5V
5.5V
+15V or 5V too low, short circuit at
preamplifier, IF board or MSB
defective
Err252 REF5V voltage shorted 24 120 +15V or REF5V too low, short
circuit at preamplifier, IF board or
MSB defective
3xx detection system (offset preamplifier, preamplifier test, emission, cathode test)
Wrn300 Anode voltage too low 41 132 7V < the
setpoint
Short circuit anode voltage,
pressure in mass spectrometer
too high, IF board, MSB or ion
source defective
Wrn301 Anode voltage too high 40 131 7V > the
setpoint
MSB defective
Wrn302 Suppressor voltage too low 39 130 297V Short circuit suppressor, IF board
or MSB defective
Wrn303 Suppressor voltage too
high
38 129 363V MSB defective
Wrn304 Anode-cathode voltage too
low
36 127 40V Short circuit anode-cathode, IF
board or MSB defective
Wrn305 Anode-cathode voltage too
high
35 126 140V MSB defective