INFICON Warning and error messages | 7
XL3000flex-Operating-instructions-jina83en1-07-(2105) 85 / 108
Err541 Sniffer blocked (p1) 62 146 Sniffer blocked, sniffer valve
defective (pressure lower than half
of the configured warning value),
filter clogged
Wrn542 Sniffer broken 64 147 Sniffer broken
Wrn550 Pressure too low, XL
Sniffer blocked
Clean or replace the high flow
capillary of the sniffer line.
Replace soiled filter.
Wrn552 XL Sniffer broken Replace the high flow capillary of
the sniffer line.
Wrn554 XL Sniffer P2 too small 63 62 Pressure on SL3000XL too low in
low flow.
6xx Calibration errors
Wrn600 Calibration factor too low 81 153 0.01 Calibration leak or machine factor
set incorrectly
Wrn601 Calibration factor too high 81 153 10000 Calibrated leak or machine factor
set incorrectly, partial flow factor
too high
Wrn602 KalFaktor lower than last
calibration
81 153 < 50% of
the old
value
Calibrated leak, machine factor or
partial flow factor has changed
Wrn603 KalFaktor higher than last
calibration
81 153 > 200% of
the old
value
Calibrated leak, machine factor or
partial flow factor has changed
Wrn604 Int. Cal. not possible, lack
of calibration leak control
81 153 calibration leak is not enabled
Wrn605 Difference during
calibration too small
Calibration leak defective or signal
too weak.
Wrn610 Machine factor too low 81 153 1.00E-04 Machine factor adjustment
inaccurate
Wrn611 Machine factor too high 81 153 1.00E+04 Machine factor adjustment
inaccurate, partial flow factor too
high
Wrn612 Machine factor lower than
last time
81 153 < 50% of
the old
value
Partial flow factor has changed
Wrn613 Machine factor greater
than last time
81 153 > 200% of
the old
value
Partial flow factor has changed