7 | Warning and error messages INFICON
86 / 108 XL3000flex-Operating-instructions-jina83en1-07-(2105)
Wrn625 Int. calibration leak not set 0 0 Leak rate of int. calibration leak is
still set to factory setting
Wrn626 Ext. Calibration leak not
set
0 0 Leak rate of calibration leak is still
set to factory setting
Wrn630 Calibration request 0 0 Temperature change of 5°C,
Rotation speed was changed
since last calibration, 30-minute
switch-on time and still no
calibration conducted
Wrn650 Calibration is not
recommended in the first
20 minutes
A calibration during the first 20
minutes after starting (warm-up
phase) the leak detector is not
recommended.
The warning message can be
turned off:
- LD protocol: Bef 429
– ASCII *CONFig:CALWarn
(ON,OFF)
Wrn670 Calibration error 81 153 Since a problem has occurred
during the calibration, you have to
recalibrate.
Wrn671 Peak not found 81 153 The signal was too restless during
the peak search. Calibration has
been aborted.
Wrn680 Deviation to the calibration
detected
0 0 The verification of calibration has
shown that you should recalibrate.
7xx temperature errors (preamplifier, electronics)
Wrn700 Preamplifier temp. too low 33 60 2°C Temperature too low
Wrn702 Preamplifier temp. too high 32 124 60°C Temperature too high
Wrn710 MSB temperature too high 54 44 58°C Temperature too high
Err711 Max. MSB temperature
exceeded
54 44 65°C Temperature too high
8xx not used
9xx maintenance messages (e.g. TMP)
Wrn901 Maintenance bearing/
lubricant
99 99 3 years TMP maintenance necessary
Wrn910 Maintenance diaphragm
pump
99 99 8000 hour maintenance of
diaphragm pump required