Weight Calibration
Upon completing Steps 1 through 7, the scale is ready to be calibrated. To ensure an
accurate calibration, one to ten load weights will be required to load the scale. Using
multiple point calibration allows the unit to weigh more accurately by removing
undesirable characteristics of load cells. A typical weight calibration is a three point
calibration. Three different and optimal loads are applied and entered (not including the
zero point). If three different weights are not available, a one or two point calibration
process may be substituted. To calibrate using one point, turn off the scale when the
screen displays the setting "". To calibrate using two points, turn off the scale
when the screen displays setting "". The CS750 can apply and load up to 10
calibration points.
8. The screen will display "" for 1 second and advance to show "". Press the
Mode button. The screen will display the current capacity setting with the last
number on the right in flashing mode. The flashing digit may be incremented by
pressing the Lb/Kg button or decremented by pressing the ZERO button. To move
one digit to the left, press the TARE button. Enter the capacity of the scale and
press the Mode button when the scale capacity is displayed.
9. The screen displays "". Remove weight from scale and press the Mode
button.
10. The screen will display setting "", enter first weight. Press the Mode button.
The screen will show "" with the right digit in flashing mode. The flashing digit
may be incremented by pressing the Lb/Kg button or decremented by pressing the
ZERO button. To move one digit to the left, press the TARE button. Enter the value
of the first load and press the Mode button when the value is displayed.
11. The screen will display setting "", load first weight. Apply first load to the
scale. When the first load is applied press Mode button.
12. The screen will display "", enter second weight. Press the Mode button. The
screen will show "" with the right digit in flashing mode. The flashing digit may
be incremented by pressing the Lb/Kg button or decremented by pressing the
ZERO button. To move one digit to the left, press the TARE button. Enter the value
of the second load and press the Mode button when the value is displayed.
13. The screen will display the setting "", load second weight. Apply second load
to the scale. When the second load is applied press Mode button.
14. The screen will display the setting "", enter third weight. Press the Mode
button. The screen will show "" with the right digit in flashing mode. The
flashing digit may be incremented by pressing the Lb/Kg button or decremented by
pressing the ZERO button. To move one digit to the left, press the TARE button.
Enter value of the third load and press the Mode button when the value is
displayed.
15. The screen will display setting "", load third weight. Apply third load to the
scale. When the third load is applied press Mode button.