CS750 User Manual
Rev E, April 2018
Page 30 of 31
Calibration Menu Skip ""
Cal Mode Number
Enter Cal Mode Number After "" in Cal Menu
000
Advance through normal calibration menu.
001
Jump to weight calibration
005
Enter Mode menu. (the same menu that is entered if the cal strap is in the
Run position)
121
Raw Counts display diagnostic
131
Constant power to all load cells diagnostic
311
Default and save all radio settings to the radio
711
Default and save all settings (Leave calibration untouched)
911
Default and save all board memory (settings and calibration)