CS750 User Manual
Rev E, April 2018
Page 21 of 31
Setting the Calibration Parameters
1. During the calibration process, data will be retained by the scale at the step
completed if the power is cycled off. To initiate calibration press the Mode button.
The screen will display "". Press the Mode button. The scale will display ""
with the right hand number flashing. To access the calibration parameters, press
Mode button once again. To skip directly to the weight calibration parameters,
proceed to step 8 and enter "". Refer to the "Troubleshooting" section for
additional codes. The flashing digit may be incremented by pressing the Lb/Kg
button or decremented by pressing the ZERO button. To move one digit to the left,
press the TARE button. When the desired number is displayed, press the Mode
button.
2. The screen will display ".". Press the mode button. The screen will show "",
indicating the lb/kg unit switch is enabled. To leave the unit switch function enabled
press the mode button. To disable the lb/kg unit function, press the Lb/Kg or ZERO
button to toggle the display to "". Press the mode button.
3. The screen will display "". Press the mode button. The screen will show the
current Motion Detect setting. When the motion detect function is enabled, any
motion detected of the load will disable the scale zero, tare, print and accum
functions. Press the Lb/Kg or the ZERO button to toggle through the motion detect
settings. When the desired setting is displayed, press the Mode button.
4. The screen will display ".". Press the mode button. The screen will show the
current ADC setting in flashing mode. This function enables the internal A/D
conversion time to either full or reduced conversion. Enter "0" to enable full
conversion. Full conversion provides the most stable results. Enter "1" to enable
reduced conversion time. Reduced conversion may extend battery life. Intercomp
recommends the conversion setting to be set to "0". It should be noted that when
the conversion setting is changed, the scale must be recalibrated. The flashing digit
may be incremented by pressing the Lb/Kg button or decremented by pressing the
ZERO button. When the desired ADC rate is selected press the Mode button.
5. The screen displays "". Press the mode button. The screen shows the current
Auto Zero Tracking (AZT) setting. Press either the Lb/Kg or the ZERO button to
toggle through the auto zero tracking options (, , ..  or .). If
after a period of time the displayed weight is less than the number of grads shown,
the weight will be zeroed off. When the correct auto zero tracking setting is
selected, press the Mode button.