Invacare®Platinum®Mobile
4Service
4.1ServicingGuidelines
WARNING!
RiskofInjuryorDamage
Impropermaintenanceandservicingmayleadto
injuryordamage.Toavoidinjuryordamage:
–UseInvacareservicepartsonly.
–Onlyprofessionalsofthehealthcareeld
orpersonsfullyconversantwiththisprocess
suchasfactorytrainedpersonnelshould
performservice,preventivemaintenance
orperformanceadjustmentsontheoxygen
concentratoranditsequipment.
–Disconnectallpowersourcespriorto
performingmaintenanceorservice,unless
otherwisedirectedbythisservicemanual.
–Thepowersupply,batterypack,andbattery
chargerarenotserviceableitems.DONOT
attempttodisassemble,modifyorservice
theseitems.
–DONOTheatthebatterypackabove140°F
(60°C).
Properlycleanalltoolspriortoperformingmaintenance
orservice.
4.2ServiceandReplacementParts
Refertothepartscatalogatwww.invacare.comfor
serviceandreplacementpartinformation.
4.3ServiceLife
WARNING!
RiskofInjuryorDamage
Useoftheproductbeyondthistimeperiodmay
causeproductdamageandinjury.
–Thisproducthasanexpectedservicelifeof
three(3)yearswhenusedinaccordancewith
safetyinstructions,maintenanceintervalsand
correctusestatedinthismanual.
–Performallmaintenanceaccordingtothe
recommendedscheduleinthismanual.
Theeffectiveservicelifecanvaryaccordingtothe
frequencyandintensityofuse.
4.4ReplacingPatientOutletFilter
1.Ensureunitisoff.
2.Disconnectanyexternalpowersourceandremovethe
batterypack(s).
3.Removetheconcentratorfromcarrybag.
4.CleanT-20driverwithalcoholpad.
5.InsertT-20driverintoopeningofoxygenoutletportI
andturncounter-clockwiseuntilbarbedcapJcomes
freefromthelowerhalfoftheport.
Filtermaylodgeinthebarbedcapafter
disassembling.
6.RemoveoldpatientoutletlterK.
7.Cleanallexposedsurfacesofbothhalvesoftheoxygen
outletportLwithacleanalcoholpad.
181187375-B