Maintenance
6.4CleaningandDisinfectionBetweenPatients
Cleaninganddisinfectionmustbeperformedbetween
patientsandasnecessary.
WARNING!
RiskofInjuryorDamage
Topreventinjuryfrominfectionordamageto
concentrator:
–Onlyqualiedpersonnelshouldperform
cleaninganddisinfectionoftheoxygen
concentratorandaccessoriesbetweenpatients.
Followtheseinstructionstoeliminatepossible
pathogenexchangebetweenpatientsdueto
contaminationofcomponentsoraccessories.
1.Disposeofandreplaceallpatientsideaccessoriesnot
suitableformultiplepatientuse,includingbutnot
limitedto:cannulaandoxygentubing.
2.Cleanconcentratorandaccessoriesasdescribedin
Maintenancesectionoftheusermanual.
3.Disinfectthesurfacesoftheconcentratorand
accessoriesusingClorox®4inOneDisinfectant&
Sanitizerorsubstantiallyequivalentproduct.
WARNING!
RiskofInjuryorDamage
–Followthecleaning/disinfectantproduct
manufacturer’sinstructionsforuse,
handling,storageanddisposalofthe
product.
–Useofinappropriatedisinfectingproducts
maydegradedevicematerialsand
negativelyaffectdevicesafetyand
performance.
Iftheaccessories,suchasthecarrybag,are
unabletobecleanedordisinfected,discardthe
accessoryandreplace.
4.PerformthetasksonPreventiveMaintenanceChecklist.
5.Beforerepackaginganddistributiontonewpatient,
ensurepackagingcontentscontaintheconcentrator,
battery,carrybagwithshoulderstraps,ACpower
adapter,ACpowercord,DCpowercable,labelsand
usermanual.
1187375-B43