Troubleshooting
General
Alarm
Name
Alarm
Code
AlarmCodeNameDisplayMessagePossibleReasonSolution
LowBattery
(~30
minutes
left)
0x7207
LPA_lowBattery30MinLeftThedisplayshows:
"LowBatteryCharge
Unit"
1.Theconcentratoris
runningoffbattery
powerandthe
batterycharge
leveliscalculated
toonlybeableto
providebetween
15to30minutes
ofadditional
operationbefore
reaching0%state
ofcharge.
1.Supplyunitwith
externalpower
(AC/DC)orinstall
achargedbattery
intotheunit.
2.Ifalarmpersists,
contactInvacare
forservice.
LowBattery
(~15
minutes
left)
0x2200
LPA_lowBattery15MinLeftThedisplayshows:
"LowBatteryCharge
Now"
1.Theconcentrator
isrunningoff
batterypower
andthebattery
chargelevelis
calculatedtoonly
beabletoprovide
15minutesor
lessofadditional
operationbefore
reaching0%state
ofcharge.
1.Supplyunitwith
externalpower
(AC/DC)orinstall
achargedbattery
intotheunit.
2.Ifalarmpersists,
contactInvacare
forservice.
1187375-B25