Troubleshooting
General
Alarm
Name
Alarm
Code
AlarmCodeNameDisplayMessagePossibleReasonSolution
Stuck
Button
0x1107
SDA_stuckButtonThedisplayshows:
"ShutDownStuck
Button"
1.Oneofthe
buttonsonthe
concentrator's
controlpanelwas
pressedwhena
powersourceis
beingconnected
totheunit.
1.Makesureno
controlpanel
buttonsare
pressed.
2.Disconnectall
powersources
thenreconnectto
apowersource.
3.Ifalarmpersists,
contactInvacare
forservice.
NoBreath
Detected
(shutdown)
0x110c
SDA_noBreathDetectedThedisplayshows:
"ShutDownNoBreath
Detected"
1.Aninhalationhas
notbeendetected
bytheunitfor
morethan120
seconds.
1.Verfycannulais
connected,not
kinked,properly
positionedand
userisbreathing
throughhis/her
nose.
2.Checktheoutlet
lter/Replacelter .
3.Ifalarmpersists,
contactInvacare
forservice.
Internal
Fault
0x1109
ricEventQueueFullAlarmThedisplayshows:
"ShutDownInternal
Fault"
1.PCBorsoftware
fault.
1.Ifalarmpersists,
contactInvacare
forservice.