InvacarPlatinum®Mobile
General
Alarm
Name
Alarm
Code
AlarmCodeNameDisplayMessagePossibleReasonSolution
Internal
Fault
0x1120
highPressureAlarmThedisplayshows:
"ShutDownInternal
Fault"
1.Blockageinoutlet
gasstream.
Blockagemay
beinternalor
external.
1.Removeanykinks
orblockagein
cannula.
2.Replacethe
patientoutletlter .
3.Ifalarmpersists,
contactInvacare
forservice.
Internal
Fault
0x1121
lowPressureAlarmThedisplayshows:
"ShutDownInternal
Fault"
1.Internalleakageof
pneumaticcircuit.
2.Blockageofinlet
ow.
1.Cleanairintake
lterscreen.
2.Ifalarmpersists,
contactInvacare
forservice.
Internal
Fault
0x1122
compressorRunningAlarmThedisplayshows:
"ShutDownInternal
Fault"
1.Compressoror
compressor's
wireharnessis
damaged.-or-
2.Thecompressor
wireharnessis
damagedorloose.
1.Ifalarmpersists,
contactInvacare
forservice.
Internal
Fault
0x1123
compressorNotRun-
ningAlarm
Thedisplayshows:
"ShutDownInternal
Fault"
1.Compressoror
compressor's
wireharnessis
damaged.-or-
2.Thecompressor
wireharnessis
damagedorloose.
1.Ifalarmpersists,
contactInvacare
forservice.
361187375-B