5.5ActivatingaMechanicalEmergencyRelease
Therearetwotypesofmechanicalemergencyrelease
primaryandsecondary.
5.5.1PrimaryEmergencyRelease
C
A
B
1.InsertapenintotheholelabeledEmergencyUpAor
EmergencyDownBonthecontrolboxC.
Penmustremaininpositionwithintheholeforthe
EmergencyReleasetofunction.Removalwillstop
thelowering.
5.5.2SecondaryEmergencyRelease
Itisrecommendedthattheprimaryemergency
releasebeused.Thesecondaryemergencyrelease
isonlyaback-uptotheprimaryemergencyrelease.
Incaseswheretheprimaryreleaseiseithernotfunctioning
orunreachable,asecondaryemergencyreleasemaybe
used.
1.PullupontheEMERGENCYgripAandpushdownon
theliftarmsBatthesametime.