Technicaldata
9.4.1ElectromagneticCompatibility(EMC)
Guidanceandmanufacturer´sdeclaration–electromagneticemission
Thepatientliftisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthepatient
liftshouldassurethatitisusedinsuchanenvironment.
Emissionstest
ComplianceElectromagneticenvironment-guidance
RFemissions
CISPR11(partly)
GroupI
ThepatientliftusesRFenergyonlyforitsinternalfunction.Therefore,itsRF
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearby
electronicequipment.
RFemissions
CISPR11(partly)
ClassB
Thepatientliftissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Harmonic
emissions
IEC61000-3-2
ClassA
Voltageuctuations
/ickeremissions
IEC61000-3-3
Complies
Guidanceandmanufacturer´sdeclaration–electromagneticimmunity
Thepatientliftisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthepatient
liftshouldassurethatitisusedinsuchanenvironment.
Immunitytest
IEC60601testlevelCompliancelevelElectromagneticenvironment–guidance
Electrostatic
Discharge(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.