EasyManua.ls Logo

Invacare ROZE - Electromagnetic Compatibility (EMC)

Invacare ROZE
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technicaldata
9.4.1ElectromagneticCompatibility(EMC)
Guidanceandmanufacturer´sdeclarationelectromagneticemission
Thepatientliftisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthepatient
liftshouldassurethatitisusedinsuchanenvironment.
Emissionstest
ComplianceElectromagneticenvironment-guidance
RFemissions
CISPR11(partly)
GroupI
ThepatientliftusesRFenergyonlyforitsinternalfunction.Therefore,itsRF
emissionsareverylowandarenotlikelytocauseanyinterferenceinnearby
electronicequipment.
RFemissions
CISPR11(partly)
ClassB
Thepatientliftissuitableforuseinallestablishmentsincludingdomestic
establishmentsandthosedirectlyconnectedtothepubliclow-voltagepower
supplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Harmonic
emissions
IEC61000-3-2
ClassA
Voltageuctuations
/ickeremissions
IEC61000-3-3
Complies
Guidanceandmanufacturer´sdeclarationelectromagneticimmunity
Thepatientliftisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerortheuserofthepatient
liftshouldassurethatitisusedinsuchanenvironment.
Immunitytest
IEC60601testlevelCompliancelevelElectromagneticenvironmentguidance
Electrostatic
Discharge(ESD)
IEC61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floorsshouldbewood,concreteorceramictile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
1195052-B61