| Appendix
148
YLS-K
12.1.2 Messages of the Alarms tab.
Type and number of alarms
The type and number of the alarms varies depending
on the configuration of the laser.
Message
E-Stop
Type
Alarm
Reset
Automatic
Description
The alarm E-Stop is activated as soon as an internal or external E-Stop
has been triggered. The E-Stop alarm usually appears in connection
with another alarm that represents the actual cause for the E-stop. The
indicator light (Interlock) on the front side of the laser lights up yellow
(Operating elements [}31]).
Possible
causes
Another alarm is output that is responsible for the E-Stop.
Fault diagnos-
tics / trou-
bleshooting
Correct the cause of the alarm / fault.
The alarm E-Stop will then be automatically reset.
Message
E-Stop button
Type
Alarm
Resetting
Automatic
Description
The alarm E-Stop button is activated when the E-Stop button on the
front side of the laser cabinet is pressed. Pressing the E-Stop button
opens the internal safety circuit. The alarm leads to a shutdown of the
laser power supply and therefore the laser emission. The indicator light
(Interlock) on the front side of the laser lights up yellow.
Possible
causes
E-Stop button is pressed.
Analysis / trou-
bleshooting
ð Unlock the E-Stop button by turning 45° clockwise.
The alarm E-Stop button will then be automatically reset.
Message
External E-Stop
Type
Alarm
Reset
Automatic