System-Related Functions
Copyright ©ITECH Electronic Co., Ltd. 64
changes, and the voltage change
precision is 100mV.
(This configuration is displayed only
when the IO pin is set to Trigger1-out)
Can not be set in load mode
On/Off: When On is selected, the List
generates a trigger signal and outputs a
trigger signal.
(This configuration is displayed only
when the IO pin is set to Trigger1-out)
Function setting of pin 7
⚫ Trigger1-out
⚫ Trigger1-in
⚫ Input
⚫ Output
On/Off: When On is selected, a trigger
signal is output when AC amplitude
changes. The accuracy of voltage change
is 100mV and is not restricted by phase.
(This configuration is displayed only
when the IO pin is set to Trigger2-out)
On/Off: When On is selected, a trigger
signal is output when the DC amplitude
changes, and the voltage change
precision is 100mV.
(This configuration is displayed only
when the IO pin is set to Trigger2-out)
Can not be set in load mode
On/Off: When On is selected, the List
generates a trigger signal and outputs a
trigger signal.
(This configuration is displayed only
when the IO pin is set to Trigger2-out)
Display the instrument model.
Display the serial number.
Display the control board version.
6.1.1 Menu function
Set OFF Mode Status
This parameter is used to set the status after the load input is off.
⚫ Select to High-Z, After the load input is Off, the dc impedance between the
terminals of the load is high, and the resistance value varies with different
models.
⚫ Select to Open-Z, After the load input is Off, the load is in open mode, and
the internal circuit and external load of the load are disconnected by relay.
Set the keyboard sound
This item can set the key sound state. If in ON mode, then when you press a
button, the load will beep. If in OFF mode, the beeper will not make a sound.