Operations introduction 
  Copyright © Itech Electronic Co., Ltd.  40 
 
 
Fig 4-4 CW mode 
Ranges 
You  can  select  the  lower  or  higher range  for CW  mode  too.  Power  can be 
edited in either of the two ranges. Low range will supply higher accuracy and 
better resolution when you set lower power. Any value you set is outside the 
maximum value of the LOW RANGE, you should select HIGH RANGE. 
Immediate power level 
You could set the power level via front panel. If the load is in CW mode, the new 
setting power level immediately changes the input at a rate determined by the 
slew rate. If the load is not in CW mode, the setting power level will be saved for 
use until switch to CR mode. 
Limited current value 
Set limited current value under CW mode. 
Transient power level 
Set A/B transient power level on front panel or by remote operation, the load 
can  continuously  toggle  between the two  levels  when  transient  operation is 
turned on. 
4.3.5 Compound Operation Mode 
IT8700P series electronic load has four compound operation modes: CV+CC, 
CV+CR(CR-LED),  CR+CC  and  CW+CC.  These  four  compound  operation 
modes add A-Limit ( limited current value) setting function under CV/CR/CW 
mode, which can help engineers solve the transient surge current problems, 
avoid to trigger the protection function and protect DUT from burning. 
⚫  The CV+CC mode can be applied to the load simulation battery and test 
the  charging  station  or  the  car  charger.  When  the  CV  is  working,  the 
maximum loading current is limited. 
 
 
⚫  The  CR+CC  mode  is  commonly  used  in  the  testing  of  voltage  limiting, 
current  limiting  characteristics,  constant  voltage  accuracy,  and  constant