Operations introduction 
  Copyright © Itech Electronic Co., Ltd.  43 
time period 
⚫  Toggled:  Generates  a  repetitive  pulse  stream  that  toggles  between  two 
load levels. Similar to continuous mode except that the transient points are 
controlled by explicit triggers instead of an internal transient generator. 
4.4.1 Continuous 
In CC mode, transient testing can be used to check the stability of the source 
voltage.  Transient  function  has  two  current  levels  (A  level,  B  level),  which 
should be in the same range (high range or low range). You can set the A/B 
level delay time and the rise/fall slew via the main frame keyboard. The slew 
rate determines the rate which the level changes at. Press    key, the load 
will  continuously  switch  between  the  A/B  levels  preset.  Transient  load  are 
usually  used  to  test  the  power  supply’s  performance  under  load  continuous 
changing  conditions.  Figure  4-6  shows  the  current  waveform  of  continuous 
transient operation mode. 
 
 
Fig 4-6 continuous transient operation current waveform 
4.4.2 Pulsed 
In  pulsed  mode,  you  can  set  A/B  value  via  mainframe  keyboard  or  remote 
control. A/B rise/fall rate and delay time should be the same.  The electronic 
load will automatically switch to A level, after maintaining A width time, switch to 
B level, it won’t switch to A level until the instrument receive the pulse signal. 
The following picture shows the current waveform in pulse transient operation. 
 
 
Fig 4-7 pulsed mode current waveform 
4.4.3 Toggled 
In toggle mode, the electronic load will switch between A level and B level when 
receiving  a  triggering  signal  after  the  transient  operation  is  enabled.  The 
following picture shows the current waveform in toggle transient operation.