Operations introduction 
  Copyright © Itech Electronic Co., Ltd.  51 
Wdc=0.00W 
3.  Press  +   to enter configuration menu 
Sync On Set 
Von 
Measure Rate 
4.  Press  to select <Von>, press    to enter. First set <Von point>, 
press    key to enter, input the programmed value, such as 1V, press 
  key to confirm 
Von    point 
Von    Latch 
Exit 
5.  Press    to select <Von Latch>,    press    key to enter,  select 
Latch to enable on/off state, such as On, press    key to confirm 
6.  Press    key to exit 
7.  Open the load’s input 
When in remote control mode (USB / RS232 / Ether-net), you can send SCPI 
command VOLT:ON <n> to set Von value; send VOLT:LATch ON to enable Von 
LATCH function. 
4.11 Protection functions 
Load has the following protection functions: over voltage protection(OVP),over 
current  protection(OCP),  over  power  protection(OPP),  over  temperature 
protection(OTP), reverse voltage alarm (LRV/RRV). 
The mainframe will act appropriately once any of the above protection is active. 
You can press any button on front panel to restore the protection function. For 
example, the electronic load come into over temperature protection, the buzzer 
will  alarm,  the  input  will  automatically  shut  down  and  Main  frame  VFD  will 
display OTP. 
4.11.1 Over voltage protection (OVP) 
If  the  OVP  circuit  has  triggered,  buzzer  alarm,  the  main  frame  screen  will 
displays (OVP), the condition will remain until they are reset. 
Operations to clear the OVP state: 
Check  whether  the  input  voltage  is  in  the  load’s  rated  voltage  or  the 
programmed  protecting  voltage  ranges.  If  it  is  outside  of  the  range,  please 
disconnect the instrument under test. Press any key on the front panel (or send 
commandPROTection:CLEar),  the  (OVP)  displayed  on  the  front  panel  will 
disappear, load exits OVP protection state. 
4.11.2 Over current protection (OCP) 
The  electronic  load  includes  both  hardware  and  software  over  current 
protection features. 
⚫  Hardware OVP: load’s maximum input current will be limited at about 110% 
of  the  current  range,  once  the  hardware  OCP  is  triggered,  the  status 
register’s  OC  bit  will  be  set;  when  the  hardware  OCP  is  removed,  the 
status  register’s  OC  bit  will  be  reset.  Hardware  over  current  protection 
won’t affect load’s on/off state. 
⚫  Software OCP: users can set load’s software OCP value, steps:  +