V02
JWA WEIGHING SCALE
43
【Error Messages】
Error Message Reasons Solutions
The CPU unable to read the
EEPROM
Return To The Manufacturer
Unable to read the 3 points
calibration range
Do the 3 points calibration
Zero Point is Too High Make sure the pan is empty when
turn on the scale or perform the 3
points calibration.
Zero Points is Too low Make sure the pan is on the scale or
perform the 3 points calibration.
Unstable Zero Point (1)Make sure there is no winds or
vibration .
(2)Change the stability /vibration
parameter accordingly.
Battery for RTC is too weak
Replace a new battery or press to
abort.
Overload (The weights on the pan
is more than +9e of the max.
capacity )
Take away the overloaded objects.
Unable to accumulate due to
contradiction in different unit . You
cannot accumulate this two
together .
Press key twice to clear the all the
accumulation or press key to
return to normal weighing mode .
The last accumulation is more than
the preset accumulation allowed .
Press key twice to clear the all the
accumulation or press key to return
to normal weighing mode .