Programming Guide 183
[ROUTe]:CLOSe
Syntax :[ROUTe]:CLOSe <space><channel_list>
Function Selects the channel list on which the GPIB processor is to operate on subsequent
commands
Example :ROUT:CLOS (@ 1:4)
:[ROUTe]:CLOSe:STATe?
Syntax [ROUTe]:CLOSe:STATe?
Function Returns the currently selected channels. For example, if channels 1 to 3, 4 to 5,
and 5 to 7 are closed, the channel list is (@1:3,4:5,5:7).
Example :CLOS (@1:2,3:5);:CLOS:STATE? returns (@1,2,3,4,5). (Do not forget the space
between CLOS and the bracket.)
Response <channel_list>
:SOURce:WAVElength:START
Syntax :SOUR:WAVE:START <space><NRf>
Function Causes the source optics module to change its internal start wavelength to the
specified value. This value is used to command the laser on the next sweep.
Example :SOUR:FREQ:START 1518.5
:SOURce: WAVElength:START?
Syntax :SOUR: WAVE:START?
Function Causes the source optics module to output its internal start wavelength used for
laser control
Example :SOUR:FREQ:START? returns 1518.5
:SOURce: WAVElength:STOP
Syntax :SOUR: WAVE:STOP <space><NRf>
Function Causes the source optics module to change its internal stop wavelength to the
specified value. This value is used to command the laser on the next sweep.
Example :SOUR:FREQ:STOP 1575.0
:SOURce: WAVElength:STOP?
Syntax :SOUR: WAVE:STOP?
Function Causes the source optics module to output its internal stop wavelength used for
laser control
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents