Commissioning
Press the key briefly for single steps, press and hold
the key to adjust the values quickly.
5.
Save the set value as the desired value by pressing the
key.
The new desired value will flash three times.
10.6 Setting the timer
Timer displays
Timer set to 15 minutes.
28 seconds until the timer ends
The timer keeps the set value for a defined period of time. The
device then stops (OFF). The timer settings can be set when the
device has been started or stopped.
The set value is saved so that it will be retained even after a
power outage.
Setting the timer
1. and simultaneously press “
t” and O. appear
on the display.
2. Changing the timer value (in minutes):
Press the key to set a higher value.
Press the key to set a lower value.
Press the key briefly for single steps, press and hold
the key to adjust the values quickly.
3. Save the set value as the desired value by pressing
the key.
The new timer value will flash three times.
The decimal point flashes until the desired value (±0.1 K) is
reached. After the desired value is reached, the timer starts and
the display alternates between the actual temperature (for 3
sec.) and the remaining time (for 2 seconds). The remaining time
< 1 min is displayed in seconds. Until the desired value is
reached, it can still be changed. The timer remains active and
will start when the new desired value is reached. If the desired
value is changed while the timer is running, the timer will be
deactivated.
After the set time elapses, a beep sounds twice and the device
goes into the
OFF state.
Stopping the timer
• Press the key for about 1 sec. -
OFF.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com