Emptying the bath tank
13 Emptying the bath tank
Danger of scalds from hot bath fluid or hot drain tap.
Please note the following when draining the bath fluid:
Hot bath fluid:
Do not drain the bath fluid when it is hot.
Environmental Hazard:
Refer to all regulations for disposing of bath fluids.
Emptying
Switch off the unit and pull the plug or disconnect the
connection to the power supply on all poles.
For baths without a drain tap, remove the circulator from the
bath tank.
Small bath tanks do not have a drain tap and can be carried for
drainging. The temperature of the bath fluid should not exceed
50°C.
Enclosed baths and refrigeration machines
Connect a suitable hose to the drain port ( 12 mm external).
Route the hose to a vessel or drain.
Open the drain valve with the knurled screw.
To reduce the weight, the bath can be partly emptied using a
hose pump (transfer pump).
Do not empty the bath in temperatures of ≤ 0°C since the
drain tap may freeze.