2. Start data logging with the and key combination.
Hold the keys down until the activation is indicated by
LO61. Data logging is also indicated by a flashing dot
after the last decimal place (
28.33.). Every record is
automatically provided with a log file name and stored in a
separate txt file.
3. The same key combination can be used to stop data
logging. Deactivation is indicated by
LO60.
Cold Mode:
You can set the refrigeration unit operating mode. Here you can switch
between automatic operation, refrigeration unit always on, and
refrigeration always off.
To do this, press the button while switching on the power switch. The
mode changes one position each time the power switch is switched on
and the button pressed. The display shows:
- „COn“
Refrigeration unit always on,
where the required cooling capacity to maintain the bath temperature is
available.
- „AUtO“
Automatic operation (factory setting)
where cooling capacity is available if needed
- „
COFF“
refrigeration unit always off,
where no cooling capacity is required.
Reading black box data:
The black box is integrated into the controller and stores all relevant
data from the last 20 minutes. In addition, persisting alarms and
warnings are logged to an alarm memory. This data can be read.
Installation is easy and is carried out step by step. Please follow the
instructions.
To do this, insert the USB flash drive into the thermostat interface and
press the “SERVICE button” + . The display then shows
“
-BB-”.