10.7 ATC - Absolute Temperature Calibration (1-point calibration)
1-point calibration:
The temperature sensor can be calibrated at any value in the
working temperature range.
Place a calibrated thermometer (resolution: 0.01°C) in the center of
the bath to measure the actual bath temperature.
50.0
50.0
48.9
CAL
1st Switch on the circulator at its mains switch
2nd Set the required calibration value as described in
"Adjusting temperature setpoint" (example: 50.0°C).
3rd Start the unit:
Press the key.
4th The bath will be heated to 50.0°C.
When the setpoint is reached, allow the bath temperature to
settle for approx. 3 minutes.
5th Start the calibration:
Press the “Service” and the key together until the comma in
the display starts to flash.
6th Read the bath temperature value on the calibrated
thermometer and round it up or down and set it.
(Example: 48.87 °C to 48.9 °C)
7. Press the key to confirm.
7th The circulator will confirm the process by displaying
"CAL".
- Err
If the value is outside a window of ±5°C, its entry will be ignored!
Error message: Err = Error