Operation of the CORIO CD via the USB interface
12 Operation of the CORIO CD via the USB interface
• PC and CORIO CD are switched on.
• Connect the CORIO CD and PC with the USB cable.
The CORIO CD reports to the PC after the installation of the driver with
the identifier "STMicroelectronics virtual COM port" COM port in Device
Manager
For comfortable operation, you can use the JULABO software EasyTemp.
There are also interface commands to the direct query available:
Input Function
STATUS current operating status
VERSION device name + voltage variant + software version
IN_SP_00 Setpoint query
IN_PV_00 current value query
IN_PV_01 Variable query
IN_PV_03 Safety sensor query
IN_PV_04 Safety potentiometer query
IN_MODE_05 Start/Stopp query
OUT_SP_00 Setpoint input
OUT_SP_01 Send setpoint
OUT_MODE_05 Start/Stop device 1/0
Remote operation
The device can be controlled remotely via the USB interface. To do this,
connect the unit to a PC via the interface.
When remote operation is active (press and hold down and
simultaneously for approximately 3 seconds) the display alternates
between the current bath temperature or “OFF” and the message “-r-”.
Switching remote operation on and off is confirmed with a short “
r
On” or “r OFF”.
Data logging
Certain characteristic parameters can be read. To do this:
1. Attach the USB flash drive.
www.GlobalTestSupply.com