Technical data
Safety precautions to IEC 61010-2-010:
Excess temperature protection, adjustable 0°C ... 170°C
Low level protection Float switch
Classification to DIN 12876-1 Class III
Alarm Optical and audible (permanent)
Ambient conditions to IEC 61010-1:
For indoor use only.
Altitude up to 200 m - normal zero.
Ambient temperature: +5 ... +40 °C
EMC requirements
The device is an ISM device of group 1 per CISPR 11 (uses HF for internal
purposes) and is classified in class A (industrial and commercial sector).
Devices of class A are intended for the use in an industrial
electromagnetic environment.
When operating in other electromagnetic environments, their
electromagnetic compatibility may be impacted.
This device is not intended for the use in living areas and cannot
guarantee adequate protection of the radio reception in such
environments.
Humidity:
Maximum relative humidity 80%, for temperatures up to 31°C,
Linear decrease to 50% relative humidity at a temperature of 40°C
Max. voltage fluctuation of ±10% are permissible.
Protection class to EN 60 529: IP 21
The device complies with Safety class I
Overvoltage category II
Pollution degree 2