Technical data
Safety precautions to IEC 61010-2-010:
Excess temperature protection, adjustable 0°C ... 170°C
Low level protection Float switch
Classification to DIN 12876-1 Class III
Alarm Optical and audible (permanent)
Ambient conditions to IEC 61010-1:
• For indoor use only.
• Altitude up to 200 m - normal zero.
• Ambient temperature: +5 ... +40 °C
EMC requirements
The device is an ISM device of group 1 per CISPR 11 (uses HF for internal
purposes) and is classified in class A (industrial and commercial sector).
• Devices of class A are intended for the use in an industrial
electromagnetic environment.
• When operating in other electromagnetic environments, their
electromagnetic compatibility may be impacted.
• This device is not intended for the use in living areas and cannot
guarantee adequate protection of the radio reception in such
environments.
Humidity:
• Maximum relative humidity 80%, for temperatures up to 31°C,
• Linear decrease to 50% relative humidity at a temperature of 40°C
• Max. voltage fluctuation of ±10% are permissible.
Protection class to EN 60 529: IP 21
The device complies with Safety class I
Overvoltage category II
Pollution degree 2