Operating manual
14
FP40-MC FP(W)50-MC
Working temperature range °C -40 ... 200 -50 ... 200
Temperature stability °C ±0,02 ±0,02
Temperature selection digital
via keypad indication on MULTI-DISPLAY (LED)
remote control via personal computer indication on monitor
Temperature indication MULTI-DISPLAY (LED)
Resolution °C 0.1
Absolute Temperature Calibration °C ±3
Temperature control PID
Heater wattage (at 230 V) kW 2,0 2,0
Cooling capacity °C +20 0 -20 -40 +20 0 -20 -40
Medium ethanol W 680 500 320 40 900 800 500 160
Refrigerant R404A R404A / R507
Electronically adj. pump capacity stages 1 ... 4
Flow rate l/min at 0 bar 11 ... 16
Pressure max. bar at 0 liter 0.22 ... 0.45
Electrical connections:
External alarm device 24-0 V DC / max. 25 mA
Computer interface RS232
Overall dimensions (WxDxH) cm 37x46x69 42x49x70
Bath opening (WxL) cm 23x14 18x12
Bath depth cm 20 15
Filling volume liters 9 ... 16 5,5 ... 8
Weight kg 48 55
Ambient temperature °C 5 ... 40 5 ... 40
Mains power connection 230 V/50 Hz V/ Hz 207-253 / 50 207-253 / 50
Current input (at 230 V) A 13 14
Current input (at 230 V) (CH) A 9+3 9+4
Mains power connection 230 V/60 Hz V/ Hz 207-253 / 60 207-253 / 60
Current input (at 230 V) A 13 14
Mains power connection 115 V/60 Hz V/ Hz ------ ------
Current input (at 115 V) A ------ ------
Mains power connection 100 V/60 Hz V/ Hz
Current input (at 100 V) A
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C Technical changes without prior notification reserved.