11.4. Status messages
Status messages Description
00 MANUAL STOP
Circulator in „OFF“ state.
01 MANUAL START
Circulator in keypad control mode.
02 REMOTE STOP
Circulator in „r OFF“ state.
03 REMOTE START
Circulator in remote control mode.
11.5. Error messages
Error messages
Description
-01 LOW LEVEL ALARM
Low liquid level alarm.
-02 REFRIGERATOR ALARM
Control cable of the refrigerated circulator or
MVS solenoid valve controller short-circuited or
interrupted.
-03 EXCESS TEMPERATURE
WARNING
High temperature warning.
-04 LOW TEMPERATURE
WARNING
Low temperature warning.
-05 WORKING SENSOR
ALARM
Working temperature sensor short-circuited or
interrupted.
-06 SENSOR DIFFERENCE
ALARM
Sensor difference alarm.
Working temperature and safety sensors report a
temperature difference of more than 25 °C.
-07 I
2
C-BUS ERROR
Internal error when reading or writing the I
2
C
bus.
-08 INVALID COMMAND
Invalid command.
-09 COMMAND NOT
ALLOWED IN CURRENT
OPERATING MODE
Invalid command in current operating mode.
-10 VALUE TOO SMALL
Entered value too small.
-11 VALUE TOO LARGE
Entered value too large.
-12 TEMPERATURE
MEASUREMENT ALARM
Error in A/D converter.
-13 WARNING : VALUE
EXCEEDS TEMPERATURE
LIMITS
Value lies outside the adjusted range for the high
and low temperature warning limits. But value is
stored.