6.4. Change-over of the warning function to shutdown function
Factory setting: Li 0
t High
High temperature limit
t Low
Low temperature limit
For the two menu items > t High < and > t Low < choose
between a warning message being signalled or a complete
shutdown of the main functional elements such as heater and
circulating pump being effected. (see page 31)
Setting >Li / 0<
An audible warning (interval tone) sounds and a meassage
appears on the MULTI-DISPLAY (LED).
t High t Low
or
Setting >Li / 1<
A complete shutdown of heater and circulating pump is
effected.
An audible alarm (continuous tone) sounds and a message
appears on the MULTI-DISPLAY (LED).
t High t Low
or .
1. Press the key until the menu item
is displayed.
2. Press enter to indicate the adjusted parameter on the
MULTI-DISPLAY (LED) (digit blinks).
3. Select the parameter with the keys (0 / 1).
4. Press enter to store the selected parameter.
Press to update the display immediately.