Maintaining the cooling performance
52
Error messages
Description
-14 EXCESS TEMPERATURE
PROTECTOR ALARM
Excess temperature protection alarm
-20 WARNING: CLEAN
CONDENSOR OR CHECK
COOLING WATER CIRCUIT
OF REFRIGERATOR
Cooling of the condenser is affected.
Clean air-cooled condenser.
Check the flow rate and cooling water
temperature on water-cooled condenser.
-21 WARNING: COMPRESSOR
STAGE 1 DOES NOT WORK
Compressor stage 1 does not work.
-22 WARNING: COMPRESSOR
STAGE 2 DOES NOT WORK
Compressor stage 2 does not work.
-23 WARNING: HIGH
TEMPERATURE ON
COMPRESSOR STAGE 1
Excess temperature on compressor stage 1.
-24 WARNING: HIGH
TEMPERATURE ON
COMPRESSOR STAGE 2
Excess temperature on compressor stage 2.
-25 REFRIGERATOR WARNING
Error in the refrigerating machine.
-30 CONFIGURATION ERROR:
CONFIRM BY PRESSING
<ENTER> ON CIRCULATOR
The configuration of the circulator does not
conform to its present use.
Press enter to automatically perform a
single modification of the configuration.
-33 SAFETY SENSOR ALARM
Excess temperature sensor short-circuited or
interrupted.
-40 NIVEAU LEVEL WARNUNG
Low liquid level warning in the internal
reservoir.
12. Maintaining the cooling performance
To maintain the full cooling performance, clean the
condenser from time to time.
Switch off the unit, disconnect mains power cable.
Hold the venting grid, pull out and remove.
Clean the ribbed condenser with a vacuum cleaner.
Replace the venting grid.
Switch on the unit.