¾"
Only for water cooled models:
Ensure circulation of cooling water by connecting the tubing to cooling
water inlet and outlet on the rear of the refrigerated circulator.
Cooling water pressure 3,5 - 6 bar.
Cooling water temperature <20 °C.
3.2. Bath fluids
Caution:
Carefully read the safety data sheet of the bath fluid used, particulary with
regard to the fire point!
If a bath fluid with a fire point of ≤65 °C is used, only supervised
operation is possible.
Recommended bath fluids:
Bath fluids Temperature range Flash point fire point
deionized water 5 °C ... 80 °C
mixture water/glycol -30 °C bis 50 °C
Thermal M +40 °C ... +170 °C >280 °C >305 °C
Thermal H +20 °C ... +250 °C >270 °C >360 °C
Thermal HY -80 °C ... +55 °C >75 °C >80 °C
Thermal H5S -50 °C ... +105 °C >110 °C >130 °C
Thermal H10S -20 °C ... +180 °C >175 °C >210 °C
Thermal H20S +0 °C ... +220 °C >230 °C >270 °C
Order No. 10 liters Bath fluids Order No. 5 liters
8 940 100 Thermal M 8 940 101
8 940 102 Thermal H 8 940 103
8 940 104 Thermal HY 8 940 105
8 940 106 Thermal H5S 8 940 107
8 940 114 Thermal H10S 8 940 115
8 940 108 ThermalH20S 8 940 109