The out commands are valid only in remote control mode.
Examples:
Command to set the working temperature >t 1< to 55,5 °C:
out_sp_00
55.5
Command to ask for the working temperature >t 1<:
in_sp_00
Response from the circulator:
55.5
11.3. List of commands
out commands: Setting parameters or temperature values.
Command Parameter Response of circulator
version None Number of software version (V X.xx)
status none Status message, error message (see page 51 )
out_mode_01 0 Use working temperature >t 1<
out_mode_01 1 Use working temperature >t 2<
out_mode_01 2 Use working temperature >t 3<
out_mode_05 0 Stop the unit = R –OFF-.
out_mode_05 1 Start the unit.
out_sp_00 xxx.xx Set working temperature. „t 1“
out_sp_01 xxx.xx Set working temperature. „t 2“
out_sp_02 xxx.xx Set working temperature. „t 3“
out_sp_03 xxx.xx Set high temperature warning limit „t High“
out_sp_04 xxx.xx Set low temperature warning limit „t Low“
out_sp_07 x Set the pump pressure stage. (1 ... 4)
out_par_06 xxx Xp control parameter of the internal controller.
out_par_07 xxx Tn control parameter of the internal controller.
out_par_08 xxx Tv control parameter of the internal controller.