MC
5
Operating manual
Description
JULABO circulators have been designed for temperature application to specific fluids in
a bath tank. The units provide pump nozzles for temperature application to an external
system (loop circuit).
PID2
ATC
3
RS232
IEC
EN
DIN
þ The circulators are operated via the splash-proof keypad. The
implemented microprocessor technology allows to set and to store
different values that can be indicated on the MULTI-DISPLAY (LED).
Three menu keys facilitate adjusting setpoints, warning and safety
functions and menu functions.
þ The PID temperature control adapts the heat supplied to the thermal
requirements of the bath.
þ Absolute Temperature Calibration (ATC3) provides a high temperature
stability in the bath. With the 3-point calibration an offset is adjusted at
three temperatures to ensure an accurate temperature pattern at the
selected spot in the bath over the full temperature range.
þ Electrical connections:
The serial interface RS232 allows modern process technology without
additional interface.
Alarm output for external alarm message or control of JULABO
refrigerating baths or solenoid valve (cooling water).
þ The excess temperature protection conforming to IEC 61010-2-010 is a
safety installation independent from the control circuit. This protection
can be indicated and set on the MULTI-DISPLAY (LED).
þ The early warning system for low level signals that bath fluid needs to
be refilled before the low level protection conforming to IEC 61010-2-
010 causes a complete shut-down of the main functional elements.
þ Pump Pressure Control. The pump capacity (electronically adjustable
via the motor speed) enables to adapt to varying conditions for internal
and external temperature applications.
þ The circulator conforms to the relevant requirements specified by
European guidelines.
JULABO circulators are not conceived for direct temperature application to
food and luxury articles or pharmaceutical and medico-technical products.
Direct temperature application means: Unprotected contact of the object
with the bath medium (bath fluid).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com