6. Safety installations, warning functions
Settings for the excess temperature protection according to
IEC 61010-2-010 and for the high and low temperature
warning functions are made
in a menu that is called up with the key .
6.1. Excess temperature protection
This safety installation is independent of the control circuit.
When the temperature of the bath fluid has reached the safety
temperature, a complete shutdown of the heater and pump is
effected.
The alarm is indicated by optical and audible signals
(continuous tone) and on the MULTI-DISPLAY (LED)
appears the error message "Error 14".
1. Press the key and call up the menu .
2. Press enter .
The adjusted cut-out value is indicated on the MULTI-
DISPLAY (LED).
3. Set the new cut-out value using a screwdriver via the
MULTI-DISPLAY (LED). (Example: 100 °C)
4. Press to update the display immediately.
Recommendation:
Set the excess temperature protection at 5 to 10 °C above the
working temperature setpoint.
Warning:
The excess temperature protection >tSA< should be set at least 25 °C
below the fire point of the bath fluid used.
In the event of wrong setting there is a fire hazard!
We disclaim all liability for damage caused by wrong settings!