EasyManua.ls Logo

Julabo F12-MC - High and Low Temperature Warning Functions

Julabo F12-MC
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MC
31
6.2. High and low temperature warning functions
Factory settings:
High temperature limit
t High 205 °C
Low temperature limit
t Low -55 °C
If for a sensitive temperature application task adherence to a
working temperature value (setpoint) is to be supervised, then
set high and low temperature warning values.
In the example below, the >Setpoint< of 85 °C is surrounded
by the values > t High < 87 °C and > t Low < 83 °C. The
electronics immediately registers when the actual temperature
attains a temperature out of the limits and it follows a reaction
according to what is set in the menu item > Li < (Warning or
Alarm).
(see chapter 6.4. Change-over of the warning function to
shutdown function).
1. Press the key until the menu item
or is displayed.
2. Press enter to indicate the adjusted value on the
MULTI-DISPLAY (LED) (digit blinks).
3. Set value:
Use the cursor keys to move left or right on the
display until the numeral you wish to change blinks.
Use the increase/decrease arrows to change the
selected numeral (-, 0, 1, 2, 3, ... 9).
4. Press enter to store the selected value.
Press to update the display immediately
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com