Preparations
20
Notice:
Please contact JULABO before using other than recommended bath fluids.
JULABO takes no responsibility for damages caused by the selection of an
unsuitable bath fluid.
Unsuitable bath fluids are liquids which e.g.
are very highly viscous
(much higher than 50 mm
2
/s at the respective working temperature)
have corrosive characteristics or
tend to cracking.
ATTENTION: The maximum permissible viscosity is 50 mm
2
/s
.
Caution:
The temperature controlling i.e. of fluids in a reactor constitutes normal
circulator practise.
We do not know which substances are contained within these vessels.
Many substances are:
inflammable, easily ignited or explosive
hazardous to health
environmentally unsafe
i.e.: dangerous
The user alone is responsible for the handling of these substances!
The following questions shall help to recognize possible dangers and to
reduce the risks to a minimum.
Are all tubes and electrical cables connected and installed?
Note:
sharp edges, hot surfaces in operation, moving machine parts, etc.
Do dangerous steams or gases arise when heating?
Is an exhaust needed when working?
What to do when a dangerous substance was spilled on or in the unit?
Before starting to work, obtain information concerning the substance
and determine the method of decontamination.